Diseño y construcción del prototipo de un dispositivo mecánico generador de flujo con presión positiva CPAP como alternativa en el tratamiento de pacientes con covid 19 en el contexto ecuatoriano

dc.contributor.advisorArroba Arroba, César Hernán
dc.contributor.authorMolina Cabrera, Katherine Victoria
dc.contributor.authorSánchez León, Christian Vicente
dc.date.accessioned2021-10-05T04:33:59Z
dc.date.available2021-10-05T04:33:59Z
dc.date.issued2021-09
dc.descriptionLa enfermedad ocasionada por COVID-19 ha provocado un impacto a nivel mundial, cambiando la rutina cotidiana a un estilo de vida completamente diferente y requiriendo una respuesta médica rápida, por esta razón, mediante el proyecto de investigación perteneciente a la DIDE-UTA denominado “CPAP SAMAY UTA – UCL” se han adaptado e innovado, por medio de ingeniería inversa, la construcción de un dispositivo de soporte ventilatorio no invasivo. Para el proceso de ingeniería inversa se partió con una réplica en CAD del modelo original desarrollado por el University College de Londres denominado “UCL VENTURA” que sirvió como patrón inicial para realizar un análisis que contempla el análisis de su funcionamiento; luego se determinó cuáles son los componentes con el objetivo de identificar aquellos que deban sufrir grandes cambios ya sean dimensionales como de materiales. Una vez identificados dichos elementos (entrada de oxígeno y soporte) mediante el análisis de diseño en software se procedió con su redimensionamiento y cálculos de momento de apriete necesario para colocar las válvulas y elementos de entrada y salida de flujo. Finalmente, mediante el mecanizado por CAM, se fabricó el prototipo y a través de pruebas de funcionamiento se obtuvo un dispositivo más económico, pequeño y de fácil manipulación, cuya presión de oxígeno a la entrada es de cuatro bares y a su salida entrega una presión positiva de hasta 20 mmH2O que ingresa a las vías respiratorias mediante una mascarilla oronasal, este prototipo cuenta con características similares al dispositivo original pero adaptado al contexto ecuatoriano.es_ES
dc.description.abstractThe disease caused by COVID-19 has caused an impact worldwide, changing the daily routine to a completely different lifestyle and requiring a rapid medical response, for this reason, through the research project belonging to the DIDE-UTA called " SAMAY UTA - UCL CPAP” have adapted and innovated, through reverse engineering, the construction of a non-invasive ventilatory support device. For the reverse engineering process, we started with a CAD replica of the original model developed by the University College of London called "UCL VENTURA" that served as an initial pattern to carry out an analysis that includes the analysis of its operation; Then it was determined which components are in order to identify those that must undergo major changes, both dimensional and material. Once these elements had been identified (oxygen inlet and support) by means of the software design analysis, they proceeded with their resizing and calculations of the necessary tightening moment to place the valves and flow inlet and outlet elements. Finally, by means of CAM machining, the prototype was manufactured and through functional tests a more economical, small and easy-to-handle device was obtained, whose oxygen pressure at the inlet is four bars and at its outlet it delivers a positive pressure of up to 20 mmH2O that enters the airways through an oronasal mask, this prototype has similar characteristics to the original device but adapted to the Ecuadorian contextes_ES
dc.identifier.urihttps://repositorio.uta.edu.ec/handle/123456789/33706
dc.language.isospaes_ES
dc.publisherUniversidad Técnica de Ambato. Facultad de Ingeniería Civil y Mecánica, Carrera de Ingeniería Mecánicaes_ES
dc.rightsopenAccesses_ES
dc.subjectCPAPes_ES
dc.subjectCOVID-19es_ES
dc.subjectVENTILACIÓN MECÁNICAes_ES
dc.subjectPRESIÓN POSITIVAes_ES
dc.subjectMASCARILLA ORONASALes_ES
dc.subjectINGENIERÍA INVERSAes_ES
dc.titleDiseño y construcción del prototipo de un dispositivo mecánico generador de flujo con presión positiva CPAP como alternativa en el tratamiento de pacientes con covid 19 en el contexto ecuatorianoes_ES
dc.typebachelorThesises_ES

Files

Original bundle

Now showing 1 - 1 of 1
No Thumbnail Available
Name:
Tesis I. M. 654 - Molina Cabrera Katherine Victoria y Sánchez León Christian Vicente.pdf
Size:
14.19 MB
Format:
Adobe Portable Document Format
Description:
TESIS A TEXTO COMPLETO

License bundle

Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: